Maximum continuous power
12 W. equivalent to 10 defib pulses of 360 J every 5 minutes
Inductance
- < 2 µH, @25 Ω
- < 3 µH, @50 Ω
- < 4 µH, @75 Ω and 100 Ω
- < 5 µH, @125 Ω
- < 6 µH, @150 Ω
- < 7 µH, @175 Ω
- < 8 µH, @200 Ω
Temperature
Operating: 10 ??C to 40 ??C (50 ??F to 104 ??F) Storage: -20 ??C to 60 ??C (-4 ??F to 140 ??F)
Humidity
10 % to 90 % non-condensing
Dimensions (WxDxH)
154 mm x 272 mm x 138.7 mm (6.07 in x 10.71 in x 5.46 in)
Weight (net)
1.54 kg (3 lb 6.2 oz)
Safety class
Complies with EN61010-1 2nd Edition, Class II product
Safety and EMC marks
CE, CSA, and C-Tick
Warranty
2-year extended warranty (no-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center)
Calibration interval
1 year
Load settings
25 Ω,
50 Ω,
75 Ω,
100 Ω,
125 Ω,
150 Ω,
175 Ω,
and 200 Ω ± 1 %
Energy (all except pulsed biphasic):
±2 % reading + 0.1 J with 25,
75 Ω through 200 Ω loads,
±1 % of reading + 0.1 J with 50 Ω load
Energy (pulsed biphasic):
±2.5 % or reading + 0.3 J with 25,
75 Ω through 200 Ω loads,
±1.5 % of reading + 0.3 J with 50 Ω load
Voltage:
±1 % of reading + 2 V with 25 Ω and 50 Ω loads,
±2 % of reading + 2 V with 75 Ω through 200 Ω loads
Current:
±2 % or reading + 0.1 A with 25 Ω load,
±1 % of reading + 0.1 A with 50 Ω through 200 Ω loads