| Flow and pressure measurements |
| Flow |
High Flow |
±300 L/min***
±1.9 %* or ±0.1 L/min** |
| Measuring bidirectional |
|
yes |
| Temperature compensated |
|
automatic |
| Pressure compensated |
|
automatic |
| Humidity compensated |
|
manually |
| Pressure |
| Flow channel |
PChannel |
-50–150 mbar
±0.75 %* or ±0.1 mbar** |
| Atmospheric pressure |
PAtmo |
500–1150 mbar
±1 %* or ±5 mbar** |
| Units |
| Flow |
|
L/min, L/s, cfm |
| Pressure |
|
inH2O, bar, mbar, cmH2O, mmHg |
| Other measurements |
Oxygen (Option)
(pressure comp. ≤ 150 mbar) |
O2 |
0–100 %
±1 % O2** |
| Gas temperature |
Temp. |
0–50°C
±1.75 %* or ±0.5 °C** |
| Gas types |
|
Air, Air/O2, O2, N2O, N2O/O2 |
| Gas standards |
|
ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD |
| Ventilation parameter |
| Breath rate |
Rate |
1–1000 bpm
±1 bpm* or ±2.5 %** |
| Time |
Ti |
0.05–60 s
±0.02 s |
| Ratio |
I:E |
1:300–300:1
±2.5 %* |
| Breath volume |
V |
±2 %* or ±0.20 mL (>6 L/min)** |
| Tidal Volume |
Vti |
±10 L
±2 %* or ±0.20 mL (>6 L/min)** |
| Minute Volume |
Vi |
0–300 L/min
±2.5 %* |
| Peak flow |
PFInsp, PFExp |
± 300 L/min
±2.0 %* or ±0.1 L/min** |
| Pressure |
PPeak, PMean, PEEP |
0–150 mbar
±0.75 %* or ±0.1 mbar** |
| General information |
| Display |
|
1.7'' with touch control elements (color), Realtime curves |
| Data storage |
|
microSD Card |
| Power |
|
100–240 VAC, 50–60 Hz |
| Dimension (w x d x h) |
|
11.4 x 7 x 6 cm |
| Weight |
|
0.38 kg |
| Battery |
|
4 hours |
| Approvals |
|
CE, BC (Energy Efficiency for Battery Charging Systems), CSA (North America), IEC 61010-1:2010, IEC 61326-2:2012 |
|
The greater tolerance is valid:
* Tolerance related to the measured value,
** Absolute tolerance,
*** The unit L/min is based on ambient conditions of 0 °C and 1013.25 mbar (DIN 1343).
|