The CITREX H4 is in all-in-one test device ideal for biomedical engineers, independent service providers, and anesthesia and respiratory equipment manufacturers. It is designed to support various daily applications. It provides accuracy and reliability for testing multiple medical devices, including respiratory and anesthesia devices, oxygen flow meters, pressure gauges, and suction devices.

Watch the videoplay_arrow

Thông số kỹ thuật
Flow and pressure measurements
Flow High Flow ±300 L/min***
±1.9 %* or ±0.1 L/min**
Measuring bidirectional   yes
Temperature compensated   automatic
Pressure compensated   automatic
Humidity compensated   manually
Pressure
High PHigh 0–10 bar
±1 %* or ±10 mbar**
Differential PDiff ±200 mbar
±0.75 %* or ±0.1 mbar**
Flow channel PChannel -50–150 mbar
±0.75 %* or ±0.1 mbar**
Atmospheric pressure PAtmo 500–1150 mbar
±1 %* oder ±5 mbar**
Units
Flow   L/min, L/s, cfm, mL/min, mL/s
Pressure   bar, mbar, cmH2O, inH2O, Torr, inHg, hPa, kPa, mmHg, PSI
Other measurements
Oxygen
(pressure comp. ≤ 150 mbar)
O2 0–100 %
±1 % O2**
Gas temperature Temp. 0–50°C
±1.75 %* or ±0.5 °C**
Gas types   Air, Air/O2, O2, N2O, N2O/O2, CO2, N2, He/O2
Gas standards   ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD, BTPD-A, 0/1013, 20/981, 15/1013, 25/991, 20/1013, NTPD, NTPS
Ventilation parameter
Breath rate Rate 1–1000 bpm
±1 bpm* or ±2.5 %**
Time Ti, Te 0.05–60 s
±0.02 s
Ratio I:E 1:300–300:1
±2.5 %*
Ti/Tcyc   0–100 %
±5 %*
Breath volume V ±2 %* or ±0.20 mL (>6 L/min)**
Tidal volume Vti, Vte ±10 L
±2 %* or ±0.20 mL (>6 L/min)**
Minute volume Vi, Ve 0–300 L/min
±2.5 %*
Peak flow PFInsp, PFExp ± 300 L/min
±1.9 %* or ±0.1 L/min**
Pressure PPeak, PMean, PEEP, PPlateau, IPAP 0–150 mbar
±0.75 %* or ±0.1 mbar**
Compliance CStat 0–1000 mL/mbar
±3 %* or ±1 mL/mbar**
General information
Realtime curves   yes
Display   1.7'' with touch control elements (color)
Interface   RS-232, USB, Ethernet, CAN, Analog Out, TTL, TSI4000 Protocol
Data storage   microSD Card
Power   100–240 VAC, 50–60 Hz
Dimension (w x d x h)   11.4 x 7 x 6 cm
Weight   0.4 kg
Battery   4 hours
Approvals   CE, BC (Energy Efficiency for Battery Charging Systems), CSA (North America), IEC 61010-1:2010, IEC 61326-2:2012