| Safety standard |
Complies with IEC 61010-1:2010, pollution degree 2 |
| EMC standard |
Complies with IEC 61326-1:2013 |
| Dimensions (w × d × h) |
90 × 37 × 154 mm (3.5 × 1.46 × 6.1 inches) |
| Weight |
400 grams (0.88 pounds) |
| Operating temperature |
+10 to +40 °C (+50 to +104 °F) |
| Operating humidity |
10 – 90 %, non-condensing |
| Storage temperature |
−20 to +50 °C (−4 to +122 °F) |
| Storage humidity |
5 – 95 %, non-condensing |
| Atmospheric pressure |
70 – 107 kPa, altitude up to 3000 m (10 000 ft) |
| Battery |
Built-in rechargeable lithium-ion. (<3600 mAh, 3.6V) |
| Battery life |
>24h continuous measurement use. |
| Charging |
5 V DC, 2 A |
| Connectivity |
USB-C for charging and PC communication |
| Display |
5-inch touch screen |
| Data storage |
> 4,000 measurements |
| Mounting |
Standard 1/4" tripod thread |
| Measurement Specifications |
| Measurement units |
mmHg, inHg, psi, mbar, bar, atm, kPa, cmH2O, inH2O |
| Barometer |
| Range |
65-110 kPa |
| Accuracy |
± 1% |
| Port 1 |
| Medium |
Dry non-corrosive gases only |
| Connector |
Male Luer-Lock |
| Range |
0 – 1200 mmHg (0 – 23 psi) |
| Accuracy |
± 0.2 % or ± 0.5 mmHg (whichever is greater)
Additional 0.02% per °C for temperatures < 20 °C and > 25 °C |
| Resolution |
0.1 mmHg |
| Pressure source range |
10 – 1200 mmHg1 |
| Pressure source accuracy |
Pressure source accuracy ± (1 % + 1 mmHg) |
| Pressure source resolution |
1 mmHg |
| Pressure relief test range |
100 – 1200 mmHg |
| Internal leak rate |
< 0.5%/min or <1 mmHg/min (whichever is greater), while connected to a 500 mL rigid volume |
| Port 2 |
| Medium |
Dry non-corrosive gases and water |
| Connector |
Miniature quick-connect plug |
| Range |
-725 – 7500 mmHg (-0.96 – 10 Bar) |
| Accuracy |
± 0.5 % or ± 2 mmHg (whichever is greater) |
| Resolution |
0.1 mmHg |
| Internal leak rate |
< 0.5%/min or <1 mmHg/min (whichever is greater), while connected to a 500 mL rigid volume |