Thiết bị khảo sát phóng xạ kiểu buồng ion nén 451P

Đảm bảo an toàn bức xạ bằng máy đo kiểm tra buồng ion hóa 451P 

Tải xuống và in Bảng dữ liệu tiếng Anh

tabs

Tổng quan
Thông số kỹ thuật
451P Radiation Detector - Pressurized Ion Chamber Survey Meter
Dimensions
4 (w) x 8 (d) x 6 in (h) (10 x 20 x 15 cm)
Weight
2.4 lb (1.07 kg)
Radiation detected
Beta above 1 MeV, Gamma and x-rays above 25 keV
Operating ranges, typical response time
  • 0 to 500 µR/h (5 µSv/h) & 5 sec
  • 0 to 5 mR/h (50 µSv/h) & 2 sec
  • 0 to 50 mR/h (500 µSv/h) & 2 sec
  • 0 to 500 mR/h (5 mSv/h) & 2 sec
  • 0 to 5 R/h (50 mSv/h) & 2 sec

  

 

 

 

 

Accuracy
Within 10% of reading between 10% and 100% of full-scale indication on any range, exclusive of energy response. Calibration source is Cs -137
Detector
Chamber: 230 cc volume pressurized air ionization chamber to 8 atmospheres or 125 psi
Controls
ON/OFF and MODE
Automatic features
Auto-zeroing, auto-ranging, and auto-backlight
 
Power requirements
Two 9 V alkaline, 200 hours operation
Warm-up time
Less than two minutes for initial operation when the instrument is in equilibrium with ambient temperature
Display LCD analog/digital with backlight
Analog
100 element bar graph 2.5 inch (6.4 cm) long. Bar graph is divided into five major segments, each labeled with the appropriate value for the range of the instrument
Digital
2.5 digit display is followed by a significant zero digit depending on the operating range of the instrument. The units of measurement are indicated on the display at all times. Digits are 0.25 inches (6.4 mm) high. Low battery and freeze indicators are also provided on the display
Modes
Integrate mode
Operates continuously 30 seconds after the instrument has been turned on. Integration is performed even if the instrument is displaying in mR/h or R/h
Freeze mode
Will place a tick mark on the bar graph display to hold on the peak displayed value. The unit will continue to read and display current radiation values
Environmental
Temperature range
- 4º to 122º F (- 20° to 50° C)
Relative humidity
0 to 100%
Geotropism
Negligible