Conductivity

4 measuring ranges - Quadropole electrode
Temperature compensation is user-selectable depending on the brand of dialysis machine
    
Range:    0 to 30.00 mS/cm
Cell constant:    typical 0,54 1/cm
Accuracy:    
0 to 199 uS/cm ± 0.6 uS/cm
200 to 1999 uS/cm ± 6 uS/cm
2 to 11.99 mS/cm ± 0.06 mS/cm
12 to 15.99 mS/cm ± 0.03 mS/cm
16 to 30 mS/cm ± 0.06 mS/cm


With option C, the conductivity/temperature probe may be disconnected. See picture above. This option is available for all meters.

Temperature

Sensor integrated in the conductivity electrode

 

Range: 0...100 °C

Resolution: 0.01 °C

Accuracy: ± 0.1 °C 

Pressure

Reading selectable as mmHg, kPa, Bar or PSI

Version HDM97BQ and BO
Range: -700 to + 1900 mmHg, 2 Bar
Resolution: 0.1 mmHg
Accuracy: 
0 ... 300 mmHg ± 1 mmHg
otherwise ± 2 mmHg
in temperature range 20 to 23 °C Ceramic sensor fluid resistant

Option H with higher accuracy
Range: -700 to + 1900 mmHg, 2 Bar
Resolution: 0.01 mmHg
Accuracy: 
0 ... 300 mmHg ± 0.5 mmHg,
1000 ... 1900 mmHg ± 2 mmHg
otherwise ± 1 mmHg
in temperature range 20 to 23 °C Sensor not fluid resistent

Flow Only Version BQ Range: 100 ... 2000 ml/min
Resolution: 1 ml/min
Accuracy: ± 2 % from reading
Power Supply Li-Ion-Polymer battery, rechargeable with 5 V DC, 500 mA
External charger:
Input Voltage 100 … 245 V / 50/60 Hz
Output Voltage 5 V DC, 500 mA
Or via USB-Interface from a computer
Operates up to 200 hours with fully charged battery
Dimension and Weight 165 x 82 x 33 mm - 6.5" x 3.2" x 1.3”
0.4 Kg - 0.88 lbs including probe