Size (HxWxL):
14.5 cm x 35 cm x 47 cm (5.75 in x 13.75 in x 18.5 in)
Weight:
7.5 kg (16.5 lbs)
Power Requirements:
100 V ac, 115 V ac, 230 V ac, 50 Hz / 60 Hz, universal input
100 V/115 V: 20 VA
230 V: 30 VA
LCD:
Monochrome 240 pixels x 64 pixels, 8 lines x 40 characters, white LED backlight
Keys :
6 (1 fixed, 5 soft-defined) and rotary selector knob
Operating temperature:
10 °C to 40 °C (50 °F to 104 °F)
Storage temperature:
-20 °C to 60 °C (-4 °F to 140 °F)
Humidity:
10 % to 90 % non-condensing
IEC 61326-1: Basic Emissions Classification:
IEC CISPR11: Group 1, Class A. Group 1 have intentionally generated and/or use conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself. Class A equipment is suitable for use in nondo¬mestic locations and/or directly connected to a low-voltage power supply network
USA (FCC):
Intentional Radiators
This device complies with part 15 of the FCC Rules. Operation is subject to the follow¬ing two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. (15.19) Changes or modifications not expressly approved by Fluke could void the user's authority to operate the equipment. (15.21)
IEC 61010-1:
Overvoltage category II, pollution degree 2
IEC 61010-2-030:
Measurement 5,000 V
FCC (United States) compliant (Class A):
FCC ID: X3ZBTMOD3
IC (Industry Canada) compliant:
IC: 8828A-MOD3
CE (European) certified:
CE0051
Measures:
Cut and coag waveforms, monopolar and bipolar outputs
Power and current measurements:
True RMS
Bandwidth:
3 MHz at -3 dB including loads – See User’s Manual Supplement for more information
Delay time for single measurements:
0.2 seconds to 4.0 seconds from Foot Switch activation to start of measurement
Variable load:
10 seconds on, 30 seconds off, at 100 W, all loads
Fixed 200 Ω load:
10 seconds on, 30 seconds off, at 400 W
Oscilloscope Output
1 V per ampere of input current, typical
Footswitch simulations
Cut and Coag
Maximum: At 25% duty cycle (10 seconds on, 30 seconds off):
10 Ω: 300 W, 20 Ω to 2900 Ω: 400 W, 3000 Ω to 5200 Ω: 200 W
At 10% duty cycle (5 seconds on, 45 seconds off):
10 Ω: 300 W, 20 Ω to 2400 Ω: 500 W, 2425 Ω to 2900 Ω: 400 W, 3000 Ω to 5200 Ω: 200 W