Range (mains voltage)
- 90 V ac rms to 132 V ac rms
- 180 V ac rms to 264 V ac rms
Range (accessible voltage)
0 V ac rms to 300 V ac rms
Accuracy
± (2 % of reading + 0.2 V)
Voltage tests
Mains and point to point
Accuracy
± (2 % of reading + 0.015 Ω)
Resistance tests
Earth resistance and point to point
Accuracy
± (5 % of reading + (2 counts or 0.2 A, whichever is greater))
Duty cycle
- 15 A to 20 A, 5 min. on/5 min. off
- 10 A to 15 A, 7 min. on/3 min. off
- 0 A to 10 A continuous
Modes*
AC + DC (True rms), AC only, DC only
* Modes are available in all leakage tests with the exception of MAP leakages that are available only in true rms
Patient load selection (input impedance)
- AAMI ES1-1993 Fig.1
- IEC 60601: Fig 15
Ranges
- 0 µA to 199.9 μA
- 200 µA to 1999 μA
- 2 mA to 10 mA
DC to 1 kHz
± (1 % of reading + (1 μA or 1 LSB, whichever is greater))
1 kHz to 100 kHz
± (2 % of reading + (1 μA or 1 LSB, whichever is greater))
1 kHz to 5 kHz (current > 1.6 mA)
± (4 % of reading + (1 µA or 1 LSD, whichever is greater))
100 kHz to 1 MHz
± (5 % of reading + (1 μA or 1 LSB, whichever is greater))
Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are:
- At 120 V ac + (2.5 µA or 1 LSD, whichever is greater)
- At 230 V ac additional ±3.0 % and + (2.5 µA or 1 LSD, whichever is greater)
For Alternative equipment, Alternative AP, and Direct AP leakage tests, the leakage values are compensated for nominal mains as per 62353. Therefore, the accuracy specified for other leakages is not applicable.
Leakage tests
- Ground wire (earth)
- Chassis (enclosure)
- Lead to ground (patient)
- Lead to lead (patient auxiliary)
- Lead isolation (mains on applied part)
- Direct equipment
- Direct applied part
- Alternative equipment
- Alternative applied part
- Point to point
Mains on applied part test voltage
- 100 % ± 7 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI
- 100 % ± 7 % of Mains for IEC 62353 current limited to 3.5 mA ± 25 % per IEC 62353
- 100 % ± 7 % of Mains for IEC 60601-1 current limited to 7.5 mA ± 25 % per IEC 60601-1
Ranges
- 75 µ A to 199 µA
- 200 µA to 1999 µA
- 2 mA to 20 mA
Accuracy
± (10 % of reading + (2 counts or 20 µA, whichever is greater))
Ranges
- 0.5 MΩ to 20 MΩ
- 20 MΩ to 100 MΩ
Accuracy
- ± (2 % of reading + 0.2 MΩ)
- ± (7.5 % of reading + 0.2 MΩ)
Source test voltage
500 V dc or 250 V dc (+ 20 %, -0 %) 2.0 ± 0.25 mA short-circuit current
Insulation resistance tests
Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part)
Accuracy
- ± 2 %
- ± 5 % for amplitude of 2 Hz Square wave only, fixed @ 1 mV Lead II configuration
Rates: ECG Complex:
30 BPM, 60 BPM, 120 BPM, 180 BPM, and 240 BPM
Square wave (50 % duty cycle)
0.125 Hz and 2 Hz
Sine wave
10 Hz, 40 Hz, 50 Hz, 60 Hz, and 100 Hz
Pulse (63 ms pulse width)
30 BPM and 60 BPM
Available Selections
ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551
IEC60601-1 2nd Edition
Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device, Generic Device and System
IEC62353
Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device
NFPA-99 (Hospital)
Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device
ANSI/AAMI ES-1
Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device
USB Device Upstream Port
Mini-B connector for control by a computer
USB Host Controller Port
Type A, 5 V output, 0.5 A max load. Connector for keyboard and barcode reader
Wireless
IEEE 802.15.4 for control by a computer
Modes of Operation
Manual and remote
Mains voltage outlet
120 V ac, 230 V ac
Mains voltage inlet power range
90 V ac rms to 132 V ac rms, 180 V ac rms to 264 V ac rms
Maximum current
20 A, 16 A
Dimensions (WxDxH)
17.6 cm x 8.4 cm x 28.5 cm (6.9 in x 3.3 in x 11.2 in)
Operating temperature
10 ºC to 40 ºC (50 ºF to 104 ºF)
Storage temperature
-20 ºC to 60 ºC (-4 ºF to 140 ºF)
Operating humidity
10 % to 90 % non-condensing
Altitude
- 120 V ac mains supply voltage up to 5000 meters
- 230 V ac mains supply voltage up to 2000 meters
Warranty
Two-year extended warranty (no-cost, available after first-year calibration at any authorized Fluke Biomedical Service Center, otherwise standard one year warranty applies)