| General Specifications |
| Safety standard compliance |
lEC 61010-1: Overvoltage category II, pollution degree 2
IEC 61010-2-034: Measurement CAT II 300 V |
| Dimensions (w x d x h) |
214 x 207 x 92 mm (8.4 x 8.1 x 3.6 inches) |
| Weight |
1.3 kg (2.7 pounds) |
| Ingress protection |
IP40 per IEC 60529, excluding equipment outlet |
| Operating temperature |
0 to +35 °C (+50 to +95 °F) |
| Operating humidity |
10-90 %, non-condensing |
| Storage temperature |
-20 to +60 °C (-4 to+ 140 °F) |
| Storage humidity |
5-95 %, non-condensing |
| Altitude |
100-127 V ac mains voltage and ≤ 150 V on input jacks: ≤ 5000 m
200-240 V ac mains voltage and ≤ 300 V on input jacks: ≤ 2000 m |
| Battery life |
Up to 2 h |
| Connectivity |
1 x USB-C for PC communication, 2 x USB-A for peripherals |
| Display |
5-inch touchscreen |
| Data storage |
> 10 000 measurements |
| Power (region dependent)1 |
90-132 V ac, 20 A MAX, 47-63 Hz
90-132 V ac & 180-264 V ac, 15 A MAX, 47-63 Hz
90-132 V ac & 180-264 V ac, 16 A MAX, 47-63 Hz |
| Note: The supply mains voltage and frequency are continuously monitored. If a deviation from specification is detected, measurements are stopped, and a message is displayed until the user confirms. |
| CSA approval |
US version,
90-132 V ac, 20 A MAX, 47-63 Hz |
CSA approved |
NEMA 6-15 version,
90-132 V ac & 180-264 V ac,
15 A MAX, 47-63 Hz |
CSA approved |
| Electromagnetic Compatibility (EMC) |
| See users manual for more information. |
| Measurement Specifications |
| Mains voltage |
| Range |
0-264 V ac rms, limited by power specification |
| Accuracy |
±(2 % + 0.2 V) |
| Point to point voltage |
| Range |
0-300 V ac rms, limited by altitude specification |
| Accuracy |
±(2 % + 0.2 V) |
| Protective earth resistance & Point to point resistance |
| Range |
0-20 Ω |
| Accuracy |
±(1 % + 0.01 Ω) at ≤ 2 Ω
±(1 % + 0.1 Ω) at >2 Ω |
| Test current |
Square wave of at least ±200 mA for ≤ 2 Ω |
| Open circuit voltage |
Max ±24 V |
| Equipment Current |
| Range |
0-20 A ac rms, limited by power specification |
| Accuracy |
±(5 % + 0.05 A) |
| Max duty cycle |
0-10 A: continuous
10-15 A: 7 min on/3 min off
15-20 A: 5 min on/5 min off |
| Note: Equipment outlet will be turned off if max duty cycle is exceeded. |
| Leakage Current |
| Modes |
ac + dc (true rms), ac only, dc only
Mains on applied part leakage test: ac only |
| Patient load |
ANSI/AAMI ES1:1993 |
| Crest factor |
≤2 |
| Range |
0 µA - 20 mA |
| Accuracy |
dc & 20 Hz - 0.5 kHz: ±(1 % + 1 µA)
0.5 kHz - 50 kHz: ±(2.5 % + 1 µA)
50 kHz - 1 MHz: ±(5 % + 1 µA) |
| Mains on applied part |
| Test voltage |
Mains voltage ±5 % |
| Current limit |
1 mA ± 25 % at 115 V for AAMI ES1, NFPA 99 |
| Additional uncertainty |
Up to ±2 µA for 120 V, up to ±4 µA for 230 V |
| Insulation Resistance |
| Test voltage 250 V dc and 500 V dc |
| Range |
0.1-100 MΩ |
| Accuracy |
±(2 % + 0.2 MΩ) at ≤ 10 MΩ
±(7.5 % + 0.2 MΩ) at >10 MΩ |
| Test voltage accuracy |
+20 % / -0 % |
|
| Test voltage 50 V dc and 100 V dc |
| Range |
0.1-20 MΩ |
| Accuracy |
±(10 % + 0.2 MΩ) |
| Test voltage accuracy |
+30 % / -0 % |
|
| Short-circuit current |
2 mA ± 0.25 mA |
| Maximum load capacitance |
2 µF |
| ECG simulation |
| Frequency accuracy |
±2 % |
| Amplitude accuracy |
±5 %, for 2 Hz square wave |
| Waveforms |
| ECG complex |
30, 60, 120, 180 and 240 BPM |
| Square wave |
0.125 and 2 Hz, 50 % duty cycle |
| Triangle wave |
2 Hz |
| Pulse, 63 ms |
30 and 60 BPM |
| Ventricular fibrillation |
|
| Respiration simulation |
| Rate |
Apnea (0 BrPM) and 10-100 BrPM in 10 BrPM steps |
| Waveforms |
Normal |
| Inspiration : expiration ratio |
1:1 |
| Impedance baseline |
1000 Ω ± 5 % between leads |
| Impedance variation(Δ) |
1 ± 0.15 Ω |
| Respiration lead |
LL or LA, user selectable |