Flow and pressure measurement
|
Flow
|
Flow-Channel
|
Flow
|
±300 L/min***
±1.65 %* or ±0.04 L/min (for 10..40 °C)**
|
Ultra-Low-Flow-Channel
|
FlowUltra-Low
|
±1 L/min***
±1.65 %* or ±0.01 L/min (for 10..40 °C)**
|
Measuring bidirectional
|
|
yes
|
Temperature compensated
|
|
automatic
|
Pressure compensated
|
|
automatic
|
Humidity compensated
|
|
automatic
|
O2 compensated
|
|
automatic
|
Pressure
|
High Pressure & Vacuum
|
PHigh
|
-1–10 bar
±1 %* or ±7 mbar**
|
Differential Pressure
|
PDiff
|
±250 mbar
±0.5 %* or ±0.1 mbar**
|
Low Differential Pressure
|
PDiff Low
|
-10–10 mbar
±1 %* or ±0.01 mbar**
|
Pressure in Flow Channel
|
PChannel
|
-50–160 mbar
±0.5 %* or ±0.1 mbar**
|
Atmospheric pressure
|
PAtmo
|
500–1240 mbar
±1 %* or ±5 mbar**
|
Units
|
Flow
|
|
L/min, L/s, cfm, mL/min, mL/s
|
Pressure
|
|
bar, mbar, cmH2O, inH2O, Torr, inHg, hPa, kPa, mmHg, PSI
|
Additional measurements
|
Oxygen
|
O2
|
0–100 %
±1 % O2**
|
Temperature
|
In Flow Channel
|
0–50°C
±1.75 %* or ±0.5 °C**
|
Dew point
|
In Flow Channel
|
-10–50°C
±2 %* or ±1 °C**
|
Humidity
|
In Flow Channel
|
0–100% RH (non condensing)
±3 % RH ** from 10% RH to 80% RH
±5 % RH ** for < 10% and > 80% RH
|
Gas types
|
|
Air, Air/O2, O2, N2O, N2O/O2, N2, He , He/O2 , CO2, customised gas types
|
Further gas types
|
CO2
|
concentration (with optional OR-703)
|
0–15 vol%, ±(0.2 vol% + 2 % of reading)
15–25 vol%, unspecified
|
N2O
|
concentration (with optional OR-703)
|
0–100 vol%, ±(2 % vol% + 2 % of reading)
|
HAL, ISO, ENF
|
concentration (with optional OR-703)
|
0–8 vol%, ±(0.15 vol% + 5 % of reading)
8–25 vol%, unspecified
|
SEV
|
concentration (with optional OR-703)
|
0–10 vol%, ±(0.15 vol% + 5 % of reading)
10–25 vol%, unspecified
|
DES
|
concentration (with optional OR-703)
|
0–22 vol%, ±(0.15 vol% + 5 % of reading)
22–25 vol%, unspecified
|
Gas standards
|
|
ATP, ATPD, ATPS, AP21, AP25, STP, STPD0, STPD20, STPD21, STPH, BTPS, BTPS-A, BTPD, BTPD-A, 0/1013, 20/981, 15/1013, 25/991, 20/1013, 23/1013, NTPD, NTPS
|
Respiratory Parameter
|
Breath rate
|
Rate
|
1–2000 bpm
±1 bpm* or ±1 %**
|
Inspiratory time
|
Ti
|
0.01–60s
±0.01 s
|
Expiratory time
|
Te
|
0.01–90s
±0.01 s
|
Inspiratory hold time
|
TiH
|
0–60s
±0.01 s
|
Expiratory hold time
|
TeH
|
0–90s
±0.01 s
|
Post-inspiratory pause
|
% Tp
|
0–100 %
±0.1 %
|
Ratio
|
I:E
|
1:300–300:1
±2 %*
|
|
Ti/TTotal
|
0–100 %
±5 %*
|
Breath Volume
|
Vti, Vte
|
±60 L
±1.75 %* or ±0.10 mL
|
Minute Volume
|
Vi, Ve
|
0–300 L/min
±1.75 %* or ±5 mL
|
Pressure
|
PPeak, PMean, PEEP, PPlateau
|
0–160 mbar
±0.75 %* or ±0.1 mbar**
|
Peak flow
|
PFInsp, PFExp
|
± 300 L/min
±1.65 %* or ±0.04 L/min**
|
Compliance
|
CStat
|
0–1000 mL/mbar
±3 %* or ±0.01 mL/mbar**
|
Trigger
|
Automatic, Adult, Pediatric, HFO, ext. Trigger
|
Adult, Pediatric, HFO; Adjustable on flow or pressure curves with user-defined limits.
|
General information
|
Power
|
|
100–240 VAC, 50/60 Hz
|
Battery operation
|
|
16 hours
|
Power consumption
|
|
Typical 5 VA, max. 25 VA (during battery charging)
|
Weight
|
|
3.2 kg
|
Dimension (w × d × H)
|
|
24 × 26 × 13 cm
|
Data Storage
|
|
Internal, USB stick
|
Display
|
|
High resolution touch-screen display 5" (800 × 480 px)
|
Interfaces
|
|
USB-A for USB stick, USB-B for FLowLab Software, individual communication, TTL for external trigger, RS232
|
Calibration
|
|
Annually
|
Operating temperature
|
|
10–40 °C (50–104 °F)
|
Operating humidity
|
|
10–90 % R.H.***
|
Approvals
|
|
CE, CSA (North America), IEC 61010-1:2010/AMD1:2016, IEC 61326-1:2020
|
The greater tolerance is valid:
* Tolerance related to the measured value,
** Absolute tolerance, with steady air flow,
*** Non-condensing,
**** The unit L/min is based on ambient conditions of 0 °C and 1013.25 mbar (DIN 1343).
|