Range (peak)
0.1 - 1500 mA
* automatically reduced for exposures longer than 3 s.
Size of Detector
14 x 22 x 79 mm (0.5 x 0.9 x 3.1 in)
Dose/dose rate
40 - 150 kVp, 1 - 14 mm Al HVL
kVp
40 - 150 kVp, up to 1 mm Cu
TF
60 - 120 kVp, up to 1 mm Cu
Range
1 nGy - 9999 Gy (0.1 μR - 9999 R)
Uncertainty
5 % or 5 nGy (0.5 μR)
Range
1 nGy/s - 500 mGy/s (5 μR/min - 3400 R/min)
Resolution
1 nGy/s (5 μR/min)
Trig level
50 nGy/s (340 μR/min)
Uncertainty
5 % or 10 nGy/s (70 μR/min) x duty cycle
Minimum dose
50 μGy (6 mR)
Minimum dose rate (peak)
10 μGy/s (70 mR/min)
Minimum dose
1 μGy (120 μR)
Minimum dose rate (peak)
0.5 μRy/s (3.5 mR/min) at > 70 kV
2.5 μRy/s (17 mR/min) at 50 kV
Minimum dose
50 μGy (6 mR)
Minimum dose rate (peak)
10 μGy/s (70 mR/min)
Uncertainty
10 % or 0.3 mm Al
Bandwidth
4 Hz - 4 kHz* (*auto adjusted based on signal level)
Minimum dose rate (peak)
0.5 μGy/s (3.5 mR/min)
Minimum dose rate (peak)
0.5 μGy/s (3.5 mR/min )
Range
1 nGy/pulse - 999 Gy/pulse (0.1 μR/pulse - 999 R/pulse)
Minimum dose rate (peak)
0.5 μGy/s (3.5 mR/min)
Bandwidth kV
0.1 - 0.4 kHz**
Bandwidth dose rate
4 Hz - 4 kHz**
* automatically reduced for exposures longer than 1.5 s
** automatically adjusted depending on signal level
Size detector
14 x 22 x 79 mm (0.5 x 0.9 x 3.1 in)
Mo/Al, W/Rh, W/Ag, W/Al, Rh/Rh, Rh/Al
20 - 50 kVp
Mo/Cu, Rh/Cu, W/Cu, W/Ti
40 - 50 kVp
W/Al
20 - 50 kVp Measuring over 40 kVp requires an X2 R/F Sensor + 2 mm Al (incl.)
Mo/Rh
32 - 40 kVp using + 2 mm Al (incl)
Mo/Cu, W/Cu, W/Ti
40 - 50 kVp, using the X2 R/F Sensor
Range
1 μGy - 9999 Gy (0.1 mR - 9999 R)
Range
10 μGy/s - 300 mGy/s (70 mR/min - 2000 R/min)
Range
18 - 50 kVp* Measuring above 40 kVp requires an X2 R/F Sensor and on W/Al +2 mm Al (incl.) (*depending on beam quality, see active compensation)
Minimum dose
50 μGy (6 mR)
Minimum dose rate (peak)
10 μGy/s (70 mR/min)
Uncertainty
2 % or 0.5 kV (without paddle) 2 % or 0.7 kV (with paddle)
Minimum dose
1 μGy (0.1 mR)
Uncertainty
5 % above 25 kV; 10% below 25 kV
Range
1 μGy/pulse – 999 Gy/pulse
(0.1 mR/pulse – 999 R/pulse)
* automatically reduced for exposures longer than 1.5 s
Size
48 x 60 x 68 mm (1.9 x 2.4 x 2.7 in)
Classification
DIN 5032 part 7 class B
Standards
Complies with relevant parts of AAPM TG18, IEC 62563-1 and IEC 61223-2-5.
Range
0.01 – 10 000 cd/m² (0.03 – 34 000 fL)
Resolution
0.001 cd/m² (0.001 fL)
Measurement area
Ø 10 mm (0.4 in)
Uncertainty illuminant A
3%
Deviation from human eye V(λ) (f1’)
<3 % (see figure Photopic Response)
Range
0.1 – 100 000 lux (0.01 – 9000 fc)
Resolution
0.01 lux (0.001 fc)
Uncertainty illuminant A
3%
Deviation from human eye V(λ) (f1’)
<3 % (see figure Photopic Response)